Scope of Application
The product is used in conjunction with adapted insulin refills and infusion lines for subcutaneous insulin infusion in diabetic patients.
Notes
Do not fit insulin-filled refills to your insulin pump during exercises.
Do not start insulin pump therapy without your doctor's supervision.
Patients should keep in touch with medical staff and adjust the basic data of the insulin pump under the guidance of the doctor.
Read the manual carefully before use and follow the instructions.
Avoid strenuous exercise during use to prevent damage to the needle and infusion line.
Check the working status of the insulin pump at any time, pay attention to the alarm, and avoid pipeline leakage, blockage, discount, etc.
If the withdrawal time exceeds 1 hour, supplement insulin according to the instructions.
Test blood sugar no less than four times a day (recommended after three meals and before going to bed), and contact medical staff in time if any abnormalities occur.
When not in use, it is recommended to turn off the Bluetooth function.
The client device should regularly check for updates and install protection software
Insulin in unopened state should be stored at 2-8°C (refrigerated), and should not be frozen.
If the activated insulin is not used up within 28 days, a new insulin cartridge also needs to be replaced to prevent the insulin cartridge from becoming invalid.
Insulin freezes at around 0°C and deteriorates at high temperatures. When outdoors in cold weather, the insulin pump must be worn close to your body and covered with warm clothing. Avoid exposing the insulin pump to ambient temperatures above 30°C, which can affect insulin activity. When the ambient temperature is too high or too low, the user must take the necessary measures to maintain insulin activity.
The maximum dose of infusion under single failure condition ≤ 2 units.
This product is intended for infusion of specified insulin only and should not be used for infusion of other fluids.
Please be sure to use the consumables in the attachment "Infusion line and insulin refill information for use", the company will try its best to assist in dealing with the problems caused by the unmatched consumables, but does not assume various legal responsibilities caused by this.
Manual
Instruction for use-InsulinPump_Rev1.0.pdf
Manufacturer
Zhengzhou Phray Technology Co., Ltd
Building A, Building 20, Zhengzhou Linkong BiomedicalPark, at the intersection of Huanghai Road andBiotechnology 2nd Street,
Zhengzhou Airport EconomicComprehensive Experimental ZoneZhengzhou HenanChina
SRNCN-MF-000039841
TEL: +86-755-27883210
Under MDR certification / assessment
PH900 Insulin Pump